Some ATR / FTIR applications require a high performance, robust, and versatile accessory to accommodate difficult materials or high temperatures. As an example, the curing of a thermoset epoxy is evaluated and results are discussed.
Some ATR / FTIR applications require a high performance, robust, and versatile accessory to accommodate difficult materials or high temperatures. As an example, the curing of a thermoset epoxy is evaluated and results are discussed.
Attenuated total reflection (ATR) has become a primary FTIR sampling technique because of its versatility and ease of use, often entailing no sample preparation. Some materials and applications, however, require an especially robust ATR accessory. Examples include
The PIKE Technologies GladiATR (Figure 1) equipped with a resistively heated plate was used in this investigation. Operational temperature of the heated plate is from ambient to 210 °C.
Figure 1
Its monolithic diamond, a scratch and fracture resistant crystal, provides a platform to withstand applied pressures of greater than 30,000 psi. This attribute is imperative to assure intimate crystal-sample contact of mechanically rigid samples, achieving the highest quality spectra. In addition, the GladiATR utilizes optimized reflective optics for maximum energy throughput, which maximizes signal to noise ratio. This, in turn, allows for higher precision for low absorbing or low concentration components within a sample.
All spectra were collected at 8 cm–1 spectral resolution co-averaging 5 scans. The two-part thermoset epoxy investigated is suitable for applications such as bonding metal, glasses, and ceramics. Epoxy components A and B were mixed per the manufacturer's specifications at room temperature, and applied directly onto the GladiATR crystal equilibrated at a prescribed temperature (80, 90, 100, 120 °C). Curing characteristics were monitored via FTIR spectra as a function of time and temperature. Once the reaction came to completion, the residual was aggressively scraped off using a scalpel without harm to the monolithic diamond due to the diamond crystal's extreme durability.
ATR is an optimal FTIR accessory for evaluating cure characteristics of adhesives. These data are useful for formulation development as well as for quality control. Comparing spectra as a function of time may be used to determine reaction kinetics by evaluating absorbance band depletion and evolution as the reaction proceeds. This method also may be used to evaluate cure time by determining the time required to reach steady-state conditions.
Figure 2
Figure 2 shows spectra collected using the GladiATR accessory. By plotting the ratio of selected absorbance bands as a function of time, reaction kinetics may be evaluated (Figure 3). At each temperature step, a brief lag period was observed followed by a decrease or increase in the value of the band ratio and eventually achieving steady-state conditions. Assuming a linear fit during the reaction phase, a kinetic constant was determined (Table I). These data exhibited an exponential increase or decay with increasing temperatures.
Figure 3
A high temperature epoxy curing study is presented as an example application where a high performance, robust ATR accessory is desirable. Through FTIR data, reaction kinetics were determined. PIKE Technologies' GladiATR provides exceptional quality data for demanding and high temperature applications.
Table I: Kinetic constant (minâ1) for curing epoxy
PIKE Technologies
6125 Cottonwood Drive, Madison, WI 53719
Tel. (608) 274-2721, Fax (608) 274-0103
Single Cell and Microplastic Analysis by ICP-MS with Automated Micro-Flow Sample Introduction
April 25th 2024Single cell ICP-MS (scICP-MS) is increasingly seen as a powerful and fast tool for the measurement of elements in individual cells, mainly due to the high sensitivity and selectivity of ICP-MS. Analysis is performed in the same way as single nanoparticle (spICP-MS) analysis, which has become a well-established technique for the analysis of nanoparticles and particles.
Hot News on Agilent LDIR, New Developments, and Future Perspective
April 25th 2024Watch this video featuring Darren Robey and Dr. Wesam Alwan from Agilent Technologies to gain insights into the future trends shaping microplastics research and the challenges of their characterization. Discover the essential components necessary for accurate microplastics analysis and learn how the Agilent 8700 LDIR system addresses these challenges. Offering rapid and precise analysis capabilities, along with easy sample preparation methods that minimize contamination, the Agilent 8700 LDIR system is at the forefront of advancing microplastics research.
Accurate and robust analysis of challenging environmental samples for regulatory compliance
April 25th 2024This application note discusses how the adverse effects of sample matrix can be circumvented with an analytical workflow that includes robust ICP-MS instrumentation offering comprehensive yet simplified analysis of samples containing high dissolved solids such as soil digests, wastewater and solid wastes. The data presented in this application note was generated over 12 hours with over 160 samples covering soil extracts and a variety of water samples.
Analysis of drinking water in compliance with EPA Method 200.8, Revision 5.4
April 25th 2024This application note describes the analytical workflow developed for the analysis of water samples over an extended period using the latest innovations in ICP-MS instrumentation. The innovations specifically highlighted in this app note are the next generation Argon Gas Dilution accessory, which simplifies the analysis of high matrix samples allowing direct analysis without prior manual diution; the HAWK consumables and maintenance assistant within the Qtegra ISDS software; the new Easy-click compact (ECC) peristaltic pump; and the new Thermo Scientific autosampler, the iSC-65 autosampler.