The simplicity of Raman spectroscopy provides an ideal method for the quantitative determination of liquid mixtures.
The simplicity of Raman spectroscopy provides an ideal method for the quantitative determination of liquid mixtures.
The use of Raman spectroscopy for routine quantitative analysis is desirable due to the simple sample handling requirements for sample analysis. Due to the many variables associated with Raman spectroscopy, specifically, laser power, Raman cross-section, wavelength and intensity stability, Raman is rarely used for quantitative analysis. Compared to Raman analysis, there are other methods that can provide a repeatable quantitative prediction of unknowns, but these methods can require extensive sample preparation or extended time for analysis. To demonstrate the capabilities of Raman spectroscopy for quantitative analysis, a two-component mixture of liquid samples was analyzed using a Raman spectrometer and the results reported.
Several mixtures of benzonitrile dissolved in ethanol were prepared at 0.1, 0.3, 0.5, 1.0, 3.0 and 5.0% concentration as standards for development of the multi-component calibration model including a 0.2% mixture to be used as an 'unknown' for analysis using the calibration model. A Jasco NRS-3100 Raman spectrometer was used to collect triplicate spectra of the standards using 532 nm excitation and a 60 s integration of the CCD with two accumulations to obtain spectra with a spectral resolution of slightly less than 4 cm–1 resolution. Samples were contained within 2 mL HPLC sample vials and a macro accessory used for excitation/collection of the sample volumes. No further correction of the Raman spectra was performed prior to development of the calibration model or analysis of 'unknown' spectra.
The various standard spectra as well as an ethanol 'blank' were used to develop the quantitative calibration model using Jasco's Spectra Manager Quantitative Analysis package. A peak ratio method using the 2234 cm–1 C-N stretch of benzonitrile vs. the 1455 cm–1 C-H bend of the ethanol was used for the development of the linear calibration curve. Initial results based on the initial triplicate spectra of each standard provided a calibration curve with a fit of 0.9998 and an R2 of 0.9996, indicating an excellent fit of the data. To evaluate the repeatability of the calibration method, additional spectral data sets of all standards were collected again after one and two weeks, providing a total of at least nine spectra of each standard for calculation of the calibration curve. The final calibration curve also provided a linear calibration curve with very little scatter, offering a fit of 0.9997 and an R2 (residual) of 0.9995 (Figure 1). As well, the analysis of the 'unknown' sample that was prepared indicated errors of less than 0.5% for the replicate analyses conducted over the two-week period.
Figure 1: Calibration model for quantitative standard mixtures, including 2nd and 3rd week standards analysis.
The results demonstrate that a reliable calibration model can be developed from Raman spectra of a mixture without further correction. The results demonstrate a correlation of 0.999 or better for the quantitative component over a range of concentrations from approximately 0.1 to 5%. Further, the Raman method provides repeatable, reproducible results and proves to be stable for extended periods.
Jasco, Inc.
8649 Commerce Dr., Easton, MD 21610
Tel. 800-333-5272; Fax 410-822-7526
Single Cell and Microplastic Analysis by ICP-MS with Automated Micro-Flow Sample Introduction
April 25th 2024Single cell ICP-MS (scICP-MS) is increasingly seen as a powerful and fast tool for the measurement of elements in individual cells, mainly due to the high sensitivity and selectivity of ICP-MS. Analysis is performed in the same way as single nanoparticle (spICP-MS) analysis, which has become a well-established technique for the analysis of nanoparticles and particles.
Hot News on Agilent LDIR, New Developments, and Future Perspective
April 25th 2024Watch this video featuring Darren Robey and Dr. Wesam Alwan from Agilent Technologies to gain insights into the future trends shaping microplastics research and the challenges of their characterization. Discover the essential components necessary for accurate microplastics analysis and learn how the Agilent 8700 LDIR system addresses these challenges. Offering rapid and precise analysis capabilities, along with easy sample preparation methods that minimize contamination, the Agilent 8700 LDIR system is at the forefront of advancing microplastics research.
Accurate and robust analysis of challenging environmental samples for regulatory compliance
April 25th 2024This application note discusses how the adverse effects of sample matrix can be circumvented with an analytical workflow that includes robust ICP-MS instrumentation offering comprehensive yet simplified analysis of samples containing high dissolved solids such as soil digests, wastewater and solid wastes. The data presented in this application note was generated over 12 hours with over 160 samples covering soil extracts and a variety of water samples.
Analysis of drinking water in compliance with EPA Method 200.8, Revision 5.4
April 25th 2024This application note describes the analytical workflow developed for the analysis of water samples over an extended period using the latest innovations in ICP-MS instrumentation. The innovations specifically highlighted in this app note are the next generation Argon Gas Dilution accessory, which simplifies the analysis of high matrix samples allowing direct analysis without prior manual diution; the HAWK consumables and maintenance assistant within the Qtegra ISDS software; the new Easy-click compact (ECC) peristaltic pump; and the new Thermo Scientific autosampler, the iSC-65 autosampler.