Reducing Background in Surface Enhanced Raman Scattering Measurements

Raster orbital scanning (ROS) is a technique that dramatically increases the signal-to-noise ratio of Raman measurements made from target materials dispersed on surface enhanced Raman scattering (SERS) substrates.

Raster orbital scanning (ROS) is a technique that dramatically increases the signal-to-noise ratio of Raman measurements made from target materials dispersed on surface enhanced Raman scattering (SERS) substrates. For SERS substrates, rastering a laser beam across the sample decreases the background and samples many more Raman-active compounds compared to traditional static-beam techniques. The lower average power created by rastering the beam with ROS dilutes the laser power over a larger sample area. This lower average power reduces the risk of damaging the substrate, which often increases the background of the Raman spectrum.

A key challenge in acquiring high quality Raman data is reducing the background component of the spectrum, which can obscure peaks and make sample identification difficult. When analyzing SERS samples, be aware that the target material likely will be dispersed non-uniformly on the substrate. As a result, the static, tightly-focused laser spot used in traditional Raman sampling will often poorly overlap the Raman-active material and measure mostly background substrate. Also, a static laser beam can cause localized heating and damage to the SERS substrate, further increasing the background of the spectrum. ROS sampling found in Ocean Optics IDRaman reader and IDRaman mini products overcomes these challenges and improves SERS measurements.

ROS Sampling of SERS Substrates

In Figure 1, Raman shift versus wavenumber is shown for a SERS-active nanoparticle membrane tagged with 2-mercaptopyridine-N-oxide measured at 4 mW of laser power. The benefit of ROS sampling is demonstrated when the spectrum measured with traditional Raman sampling (emulated by turning ROS sampling off) is compared to the spectrum measured with ROS sampling. When raster scanning is enabled, the background (represented by the hashed line in the ROS off example) is much lower and peaks are more clearly resolved. This allows simpler and more confident identification of the target material.

Figure 1: A SERS sample measured at 4 mW of laser power with and without ROS sampling. ROS reduces the background component of the spectrum by collecting more data from the Raman-active hotspots relative to the background. The elevated background of a traditional, static laser beam sampling method is illustrated by the hashed line.

Furthermore, the effect of localized heating damage on SERS substrates is demonstrated in Figure 1 via the hashed, horizontal line on the y-axis. When ROS is not used, the background increases substantially. The constant movement of the laser spot with ROS turned on (lower spectrum) minimizes sample damage caused by localized heating from a static, focused laser spot. Therefore, ROS enables the use of higher laser power without degrading the quality of the Raman measurement.

Conclusions

As shown in this application note, the measurement of SERS samples can be greatly improved with the use of ROS. ROS reduces the background component of SERS spectra and helps to reveal features previously hidden in the high background of SERS substrate measurements. In addition, ROS samples a larger area and facilitates the use of less expensive, less homogeneous SERS substrates with the same enhancement of Raman activity. Lastly, ROS enables the use of higher laser power without the risk of damaging the SERS substrates, increasing the signal enhancement realized by the SERS technique.

Ocean Optics

830 Douglas Avenue, Dunedin, FL 34698

tel. +1(727) 733-2447, fax: +1(727) 733-3962

Website: www.OceanOptics.com

Articles in this issue

almost 13 years ago

PIKE Technologies

almost 13 years ago

Amptek, Inc.

almost 13 years ago

Ocean Optics

almost 13 years ago

Photonis USA

almost 13 years ago

Shimadzu Scientific Instruments

almost 13 years ago

Milestone, Inc.

almost 13 years ago

Savillex Corporation

almost 13 years ago

WITec GmbH

almost 13 years ago

Thermo Fisher Scientific

Related to this article

Atoms Don't Lie–The Remaking of Atomic Spectroscopy, © Hallie -chronicles-stock.adobe.com
Atomic spectroscopy has quietly gone from counting elements in bulk to counting single nanoparticles, imaging tissue at the micrometer scale, and reading Martian rocks from several meters away. The instruments got smaller, the software got smarter, and the questions they can now answer got a lot more interesting.
Tutorial Tuesday: When to Use Infrared Spectroscopy
In this “Tutorial Tuesday” clip, Peter Larkin, who is a Manager of Spectroscopy, Thermal Analysis, and Chromatography at Syensqo, highlights how people use the method or technique that they are comfortable with, even if it’s not the best one to use.