Milestone's UltraWAVE microwave digestion system with its groundbreaking single reaction chamber (SRC) technology enables the simultaneous digestion of up of to 22 different food sample types with different chemistries at temperatures and pressures as high as 300 °C and 200 bar, greatly increasing efficiency and simplifying product development.
Elemental impurities are monitored in foods to ensure safe levels and compliance with regulation. One particular class of these elemental impurities, the "Big 4" (arsenic, mercury, lead and cadmium), are of special concern because ingestion of quantity relatively low in food product can cause serious health issues, including - but not limited to - reproductive troubles, developmental disorder, brain damage, kidney and liver damage, and cancer. Given the serious health risks, a more stringent regulation is calling for the use of closed vessels digestion for sample preparation and inductively coupled plasma (ICP) for elemental impurities analysis as a replacement of obsolete and subjective methods of analysis based on colorimetric tests.
In this application note, two frosting samples are digested simultaneously on a 15-position rack with a single temperature profile. In addition, setting the instrument is straightforward: a ramp time to target temperature and a hold time at target temperature.
Frosting or icing products are delicate matrices to digest because they are made of significant amounts of carbohydrate and lipids, and thus are very exothermic during degradation. Digestion of the two varieties of frosting using a traditional microwave system would require two different runs and could possibly require two different rotors. Because of the exothermic nature of the sample, on a traditional microwave there is a limitation of the size of the sample that could be digested, since a relatively large sample size (>0.500 g) will generate a level of pressure that will cause the system to vent with loss of volatile elements.
Table I: Sample and chemistry
Samples were placed in a 15-position rack and lowered into the single reaction chamber. The chamber was pre-pressurized with nitrogen at 40 bar to raise the acids boiling point and thus prevent any cross-contamination and loss of volatiles due to boiling and spitting as the temperature increases inside the chamber.
The two samples were fully digested as evidenced by the clear solutions obtained. The digestion lasted for less than 1 h including cooling time. The higher performance and greater ease of use make the UltraWAVE a good fit for a food company where samples of different types are common. Additionally, by using five position vials, samples as large as 2 g can be digested. Such capability not only addresses the common sample homogeneity concern of the food industry, but also gives the possibility to dilute samples with as much flexibility for ICP-OES as for ICP-MS.
Milestone, Inc.
25 Controls Drive, Shelton, CT 06484
tel. (866) 995-5100, fax (203) 925-4241
Website: www.milestonesci.com
Atomic Perspectives: Highlights from Recent Columns
March 3rd 2025“Atomic Perspectives,” provides tutorials and updates on new analytical atomic spectroscopy techniques in a broad range of applications, including environmental analysis, food and beverage analysis, and space exploration, to name a few. Here, we present a compilation of some of the most popular columns.
Pittcon 2025: Highlighting Talks on Atomic Spectroscopy
February 26th 2025At Pittcon this year, there will be numerous sessions dedicated to spotlighting the latest research that uses atomic spectroscopy or elemental analysis techniques. We highlight some of these talks below that might pique the interest of spectroscopists and researchers attending the conference this year.
The 2025 Emerging Leader in Atomic Spectroscopy Award
February 15th 2025Benjamin T. Manard has won the 2025 Emerging Leader in Atomic Spectroscopy Award for his pioneering research in nuclear material characterization and isotope ratio analysis, with expertise in advanced atomic spectrometry techniques such as inductively coupled plasma optical emission spectroscopy (ICP-OES), inductively coupled plasma mass spectrometry (ICP-MS), and laser ablation.
Applications of Micro X-Ray Fluorescence Spectroscopy in Food and Agricultural Products
January 25th 2025In recent years, advances in X-ray optics and detectors have enabled the commercialization of laboratory μXRF spectrometers with spot sizes of ~3 to 30 μm that are suitable for routine imaging of element localization, which was previously only available with scanning electron microscopy (SEM-EDS). This new technique opens a variety of new μXRF applications in the food and agricultural sciences, which have the potential to provide researchers with valuable data that can enhance food safety, improve product consistency, and refine our understanding of the mechanisms of elemental uptake and homeostasis in agricultural crops. This month’s column takes a more detailed look at some of those application areas.
Best of the Week: Seed Vigor, Flower Classification, Emerging Leader in Atomic Spectroscopy
January 10th 2025Top articles published this week include two peer-reviewed articles that explore optical detection technology for seed vigor and classifying flowers, as well as a profile on Benjamin Manard, who was recognized as the winner of the 2025 Emerging Leader in Atomic Spectroscopy.