Sample Preparation for NIR Analysis

Author(s)Retsch, Inc.

A much discussed issue related to near infrared analysis (NIR) is the necessity of sample preparation. Ground and unground samples of wheat were analyzed to determine how truly necessary it is to homogenize samples before NIR analysis. The new Retsch TWISTER mill, specifically designed for NIR sample preparation, was utilized for this comparative study.

A much discussed issue related to near infrared analysis (NIR) is the necessity of sample preparation. Ground and unground samples of wheat were analyzed to determine how truly necessary it is to homogenize samples before NIR analysis. The new Retsch TWISTER mill, specifically designed for NIR sample preparation, was utilized for this comparative study.

Feed quality decisively influences the productivity of livestock farming. However, high-quality forage considerably adds to costs of stock breeding. Consequently, economic considerations focus on quality and quantity of components used in the production process. This includes reliable analyses of feedstuff and ingredients from receipt of goods to final inspection as the best way to ensure a balanced feeding of livestock while keeping cost and profit orientation in focus.

Near infrared (NIR) spectroscopy is the most important analytical method for determination of protein content, moisture, fat, and ash in feeds and forage. NIR offers a quick method for analysis without the requirement of consumables or reagents. In order to utilize NIR effectively and efficiently, one must first consider the limitations of this analytical tool, especially regarding the samples to be analyzed.

Experimental Conditions

NIR analysis is sensitive to the homogeneity and analytical fineness as both of these characteristics can cause inaccurate readings due to inefficient scattering of light and detection from the analyzer. In order to demonstrate this phenomenon, samples of wheat were taken. The wheat material was divided into 10 samples of ground material and 10 samples of unground material.

The individual samples of wheat were processed in the TWISTER mill. The mill utilizes both friction and impact between the rotor and the abrasive surface of the grinding ring. The unit is designed for gentle but fast processing so that the feed material remains in the grinding chamber for a very short time, thus maintaining moisture content. The design of the mill uses a cyclone configuration that generates a vacuum jet of air that allows the sample to be pulled through the grinding chamber and collected into a sample bottle. This feature of the TWISTER allows the user to process sample-after-sample without cleaning the grinding chamber between runs.

The quantitative measurements were carried out with the NIR Spectrometer TANGO from Bruker GmbH (Billerica, Massachusetts).

Results

Test results, seen in Figure 1, illustrate the accuracy of analyzing homogenized wheat samples versus unground wheat samples. One detail to note is the significant deviations between each measurement of the unground sample, whereas, the ground wheat sample shows no significant statistical errors.

Figure 1: The analysis of the unground sample clearly shows a systematic and a considerable statistical error.

Conclusion

The general consensus for NIR sample preparation is that it is not necessary; however, from the results presented in this application study, this opinion needs revision. When considering sample preparation within the NIR analytical lab, the necessity of quick grinding processes and ease of cleaning is paramount. With Retsch's TWISTER mill, these requirements can be met, as well as the ability to provide a sample that is at analytical fineness and homogeneous. Only then is it possible to guarantee meaningful and reliable analytical results.

Retsch Inc.

74 Walker Lane, Newtown, PA 18940

tel. (866) 473-8724

Website: www.retsch-us.com


Related to this article

Infrared Reimagined: How FT-IR Spectroscopy Learned to Read Molecules, ©  Luis Eduardo  -chronicles-stock.adobe.com
FT-IR spectroscopy, long treated as a mature bench technique for confirming a carbonyl stretch or fingerprinting a polymer, has quietly become one of chemistry’s most versatile discovery engines: it now infers molecular structure straight from a spectrum without a reference library, resolves chemistry tens of nanometers wide, and screens a fingerstick of blood for disease in minutes. The last five years of published research show FT-IR moving from a confirmatory tool into a predictive, autonomous, and field-ready analytical platform.
The Ongoing Decline of Spectroscopy Expertise
In this panel discussion, five SpecAcademy instructors examined why spectroscopy expertise has declined and how SpecAcademy aims to rebuild it through curated, fundamentals-based education suited to increasingly complex analytical work.