Application Notebook
While crystalline silicon (c-Si) is still the material of choice for modern solar cell manufacturing, the drive toward lower cost, more efficient solar cells is prompting a renewed interest in thin-film amorphous silicon (a-Si). In light of this, it is more important than ever to have the ability to quickly identify Si crystallinity, both qualitatively and quantitatively. In this application note, we evaluate how a portable Raman spectrometer can be used in conjunction with a portable video microscope to measure crystallinity.
While crystalline silicon (c-Si) is still the material of choice for modern solar cell manufacturing, the drive toward lower cost, more efficient solar cells is prompting a renewed interest in thin-film amorphous silicon (a-Si). In light of this, it is more important than ever to have the ability to quickly identify Si crystallinity, both qualitatively and quantitatively. In this application note, we evaluate how a portable Raman spectrometer can be used in conjunction with a portable video microscope to measure crystallinity.
For this study, the i-Raman® from B&W Tek, was chosen due to fact that it is the highest resolution (~3 cm-1 ) portable Raman spectrometer on the market. This, in conjunction with its TE cooled CCD and convenient fiber optic probe, made it the ideal solution for both "at line" and laboratory measurement. The minimal fluorescence background of Si allowed for the use of 532 nm excitation to take advantage of the higher photon energy and shallower penetration depth as compared to longer excitation wavelengths.
In the experiment, the i-Raman® was coupled to the BAC151A video microscope from B&W Tek to facilitate accurate sample positioning for the measurement. Figure 1a shows the experiment setup of the i-Raman® and the video microscope sampling system. The sample set for this experiment contained samples with differing crystallinity, ranging from < 10% to pure crystalline Si. These Si samples were deposited on glass substrates.
Figure 1: (a) Experimental setup of the i-Raman coupled to the BAC151A video microscope, and (b) Raman spectrum of Si at varying degrees of crystallinity.
Two Raman peaks are related to silicon crytallinity: the peak at 480 cm-1 is associated to the amorphous phase and the peak at 520 cm-1 is associated with crystalline Si.
Figure 1b overlays the Raman spectra of all five samples in the set. The red and blue spectra represent samples with a crystallinity < 10%, the brown spectrum is of a sample with a crystallinity between 40% and 60%, the green spectrum is of a sample with > 90% crystallinity and the dark blue spectrum is of pure c-Si. The shifting of the Si band from 480 cm-1 to 520 cm-1 can be clearly observed in the data. As silicon crytallinity increases, the amorphous peak 480 cm-1 reduces while the microcrystalline silicon peak 520 cm-1 emerges.
Figure 1b shows how Raman spectroscopy can easily be used as a qualitative measurement to determine the difference between a-Si and c-Si, but the high resolution of the iRaman® also makes it possible for the development of calibration curves using chemometric algorithms in a software package such as B&W Tek's new BWIQ™ software package. From these curves, a method can be developed for the quantitative analysis of Si crystallinity.
B&W Tek's i-Raman® spectrometer, when coupled with a BAC151A video microscope sampling system, is an ideal setup for Si crystallinity analysis. The compact form factor and light weight (~6.6 lb) of the i-Raman when coupled with low noise TE cooled detector and high resolution F/2 spectrograph make it the ideal solution for both laboratory and field analysis.
B&W Tek, Inc.
19 Shea Way, Newark, DE 19713
tel. (302) 368-7824; fax (302) 368-7830
Website: www.bwtek.com
Testing Solutions for Metals and PFAS in Water
January 22nd 2025When it comes to water analysis, it can be challenging for labs to keep up with ever-changing testing regulations while also executing time-efficient, accurate, and risk-mitigating workflows. To ensure the safety of our water, there are a host of national and international regulators such as the US Environmental Protection Agency (EPA), World Health Organization (WHO), and the European Union (EU) that demand stringent testing methods for drinking water and wastewater. Those methods often call for fast implementation and lengthy processes, as well as high sensitivity and reliable instrumentation. This paper explains how your ICP-MS, ICP-OES, and LC-MS-MS workflows can be optimized for compliance with the latest requirements for water testing set by regulations like US EPA methods 200.8, 6010, 6020, and 537.1, along with ISO 17294-2. It will discuss the challenges faced by regulatory labs to meet requirements and present field-proven tips and tricks for simplified implementation and maximized uptime.
Practical Autodilution for ICP-MS and ICP-OES
January 20th 2025Gain insights into improving efficiency and accuracy in elemental analysis through automated dilution technology. Learn about the key capabilities of the Agilent ADS 2 system and its seamless integration with ICP-MS and ICP-OES workflows.
UV-Vis Spectroscopy: Exporting Your Measurement Out of the Instrument
January 20th 2025Optical fibers in ultraviolet-visible (UV-Vis) spectroscopy can enable measurements outside the traditional sample compartment. This paper details the components needed for fiber optic systems, such as couplers and probes, and reviews the performance of Agilent's Cary series instruments. It is crucial to choose the right fiber optic setup for a specific lab’s needs to ensure accurate and efficient measurements.