
- The Application Notebook 09-01-2021
- Volume 36
- Issue S9
Optimizing Spectral Data Using a Variable Angle ATR
Here we explore the benefits of using a variable angle attenuated total reflection (ATR) accessory to generate the most informative spectral data. Example spectra of toluene are shown.
The appearance of an ATR spectrum is influenced by several variables including the refractive index of the ATR crystal (n1) and sample (n2), the number of internal reflections, the quality of contact between the sample and the angle of incidence (AOI) of the infrared (IR) beam. Having the flexibility to change some parameters, specifically the AOI and ATR crystal type, allows optimization of spectral data to obtain the most useful information for the application.
Materials and Methods
Mid-infrared spectra of toluene were collected using the variable angle, multiple reflection ARTMax II accessory (PIKE Technologies) configured with an AMTIR trough plate. With the ATRMax II, the AOI may be adjusted by turning the center mirror assembly position dial, where the resulting effective AOI is a function of the face angle of the AMTIR crystal and the accessory position dial setting. For data presented, the face angle of the AMTIR crystal was 45o. Spectra were collected at several AOI values. 30 s of data were co-added for each measurement at a resolution of 4cm-1.
Results
Most single reflection and multiple reflection ATR accessories have a fixed nominal AOI of 45o. When using these accessories to measure samples with a refractive index of roughly 1.5, the depth of penetration (dp) is about 2 microns at 1000 cm-1 for a diamond or ZnSe ATR element and 0.6 microns for a Ge ATR element. dp decreases at shorter wavelengths and increases at longer wavelengths.
For total internal reflection to occur, the AOI must be greater than the critical angle (θ), which is defined as θ =sin-1(n2/n1). dp decreases with increasing AOI while dp increases exponentially as the angle of incidence nears the θc. By probing deeper into the sample—making measurements closer θc—the magnitude of the absorbance bands increases significantly. By using a variable angle ATR accessory such as the PIKE Technologies’ VeeMAX III single reflection or the multiple reflection ATRMax II, the absorbance band strength may be optimized to provide maximum chemical information by easily changing the AOI, whether the application requires probing more deeply into the sample by making measurements closer to θc or probing closer to the surface at grazing AOI.
Unlike a variable angle, single reflection ATR where only the dp is changing with varying the AOI, the number of reflections in addition to the dp is changing when using a variable angle, multiple reflection accessory. Figure 1 shows toluene spectra collected at 43o and 50o effective AOI where the number of reflections on the sample is 7 and 5, respectively. The spectrum collected at 43o shows absorbance bands over 50% greater than the spectrum collected at 50o. Therefore, the increase in absorbance band strength is because of two contributing factors, increased dp and the number of reflections. In addition, changing the ATR crystal material provides added flexibility to accommodate various ATR probing depths into the sample.
Conclusion
Variable angle ATR accessories can optimize spectral data for a given application. By simply changing the angle of incidence, it is possible to probe more shallow or deeper into the sample, as needed.
PIKE Technologies, Inc.
6125 Cottonwood Drive, Madison, WI 53719
Tel. (608) 274-2721, Fax (608) 274-0103
Website: www.piketech.com
Articles in this issue
over 4 years ago
Cryogenic Raman Imaging of a Tungsten Diselenide Flakeover 4 years ago
Application Note: Raman Spectroscopy for Reaction Monitoringover 4 years ago
Laser Frequency Offset Lockingover 6 years ago
A Diamond ATR Study of Worn Surfaces on MetalNewsletter
Get essential updates on the latest spectroscopy technologies, regulatory standards, and best practices—subscribe today to Spectroscopy.