Characterization of Surface Orientation of a Molded Polymer by FTIR-ATR

Thermotropic liquid crystal polymers (TLCP) are injection molded and undergo spontaneous ordering while flowing through the mold. The degree of orientation depends on the distance from the injection point, the depth into the material, and other factors. Since the orientation changes the physical properties of the product, it is important to control the degree of anisotropy. This application note demonstrates using ATR to examine the changes in surface orientation as a function of the distance from the injection point or gate.

Experimental

Samples of the TLCP were cut down the centerline of the mold, 15 mm apart, and cleaned using Alconox™ in an ultrasonic bath to remove surface contaminants.

Infrared spectra were collected on an FT-IR spectrometer equipped with the Seagull variable angle reflection accessory with its ATR rotator (Figure 1), ZnSe ATR crystal, and KRS-5 wire grid polarizer set to s-polarization. The Seagull was used at a 55° incident angle. The spectrometer collected data at 4 cm-1 resolution, signal averaged over 32 scans.

A background was collected from the clean ATR crystal. Each sample was oriented on the ATR rotator so its cut edges corresponded to the 0° and 90° setting. Sample spectra were collected at 0°, 45°, and 90°.


Results and Discussion

Figure 2 shows the data collected from the sample farthest (60 mm) from the gate. The band at 1500 cm-1 shows the most striking change. It is much weaker at the 0° position than at the 90° position. This band is due to the skeletal vibration of the benzene and naphthalene rings in the TLCP, which are oriented relative to the polymer chain axes. The band at 1730 cm-1, on the other hand, shows very little change.

Pirnia and Sung (1) developed a methodology using the dichroic ratios to calculate the orientation function for polymers and determined that the orientation was the highest on the surface, well beyond the gate. The spectra measured here were processed using the methods developed by Pirnia et al. The orientation function (Figure 3) shows the surfaces of the samples become more oriented further from the gate, comparable to the 0.77–0.88 reported.

Conclusion

The orientation of injection-molded polymers can be analyzed by ATR dichromism. Using a variable angle ATR accessory like the Seagull and its ATR rotator, the incident angle can be optimized for a given sample and the orientation axis can be determined.

Reference

1.) A. Pirnia and C.S.P. Sung, Macromol. 21(9), 2699–2706 (1988).

Harrick Scientific Products, Inc.

141 Tompkins Ave., 2nd Floor, Pleasantville, NY 10570

Tel. (800) 248-3847, fax (914) 747-7209

www.harricksci.com


Related to this article

A robotic arm in an automated agricultural setting, carefully processing a conveyer belt with tomatoes and leafy greens for food production. | Image Credit: © Donnangelo - stock.adobe.com
Croatian researchers outline how optical emission spectroscopy (OES) can track reactive species and plasma parameters in real time, giving processors a path to move cold plasma treatments from the laboratory to production lines.
Spectroscopy Top 10 DOI Articles of the Month (October 2026)
The following articles are the most accessed digital object identifier (DOI) manuscripts for Spectroscopy and LCGC International during the month of October 2026. Nine articles are individually ranked here by DOI page views; the seventh-highest entry (99 views) was recorded against the bare LCGC North America journal-level DOI rather than an individual article and is therefore not attributable to a single manuscript.
Mapping Molecular Mobility Pixel by Pixel
Garth Simpson of Purdue University explains how patterned photobleaching and spatial Fourier transforms can turn a single-point measurement into a diffusion coefficient at every pixel in the field of view.
Optical and Electronic Instrument Components,  ©  ridwan -chronicles-stock.adobe.com
The updated and expanded 2026 edition of Spectroscopy's Instrument Components Terminology Guide defines the sources, detectors, optical and dispersive elements, sampling hardware, physical constants, and unit conversions used across UV-vis, NIR, IR, Raman, X-ray, and atomic spectroscopic instrumentation, with new terms reflecting component developments reported between 2022 and 2026.