Infrared spectroscopy has long been recognized as a selective and sensitive technique for analysis and characterization of surfaces. From material identification to quantitative analysis, infrared analyses can establish if a surface is composed of the correct material, has the proper coating thickness, is free of contaminants, is properly cured, or has the correct properties for the next step in a manufacturing process.
Infrared spectroscopy has long been recognized as a selective and sensitive technique for analysis and characterization of surfaces. From material identification to quantitative analysis, infrared analyses can establish if a surface is composed of the correct material, has the proper coating thickness, is free of contaminants, is properly cured, or has the correct properties for the next step in a manufacturing process. With these analytical abilities, infrared spectroscopy could easily be used for qualification of materials and surfaces on very large objects. However, with current bench top infrared spectrometers, it is often necessary to remove a sample from a large object and bring the small piece to the spectrometer located in the laboratory. Thus, a potentially non-destructive methodology for large objects becomes a destructive one due to the necessity of small sample size coupled with the location of the FT-IR system.
Times have changed, though. Now portable, handheld FT-IR spectrometers are available which can be used for true non-destructive analysis of large surfaces. An example of this type of instrument is A2 Technologies' Exoscan™ , which can conduct non-destructive surface measurements on a variety of samples. Though size and portability is important for these applications, equally important is overall spectroscopic performance since surface measurements can be very demanding. Exoscan has a performance envelope equivalent or better than analytical bench top systems. The recent addition of interchangeable sample interfaces and a docking station provides a system that can easily be used for experimental methods and procedures in the lab, then taken to the field for routine analysis when needed.
The Exoscan (see Figure 2a) can measure surface samples with three different techniques: attenuated total reflectance (ATR), specular reflectance (reflectance) and grazing angle reflectance (grazing angle). The three interchangeable sample interfaces are shown in Figure 1. The three techniques provide a wide variety of options for many different sample types.
Figure 1
The ATR interface is ideal for the analysis of solids, liquids, pastes and gels. The interface is comprised of a diamond window, which makes it impervious to corrosion and scratching. The sample is brought into contact with the diamond window. A pressure device allows compression of the sample to ensure optimum contact for analysis. ATR is a surface technique and only the top 2–3 microns are analyzed.
The specular reflectance interface allows the analysis of films and coatings on reflective metal surfaces such as aluminum or steel. The angle of incidence is 45°. The infrared energy either passes completely through the sample before being reflected by the substrate, or the energy is reflected off the front surface of a smooth sample. One particularly useful application of the reflectance technique is determining the thickness of coatings on metal surfaces.
The grazing angle reflectance interface is similar in concept to the specular reflectance interface. It differs in the angle of incidence. The Exoscan grazing angle has a nominal angle of 82°. The high throughput optic system yields over 80% throughput, producing excellent signal to noise even on short measurements. In addition to the increased path length, grazing angle reflectance shows additional sensitivity compared to a typical specular reflectance measurement. The increase in sensitivity is due to enhancement of the electric field of the p-polarized light at the surface. The grazing angle is ideal for the analysis of very thin (sub-micron) films, such as trace contamination on reflective metal surfaces.
In addition to the multiple sample interfaces available, a docking station is also available for the Exoscan. The docking station converts the hand held Exoscan into a bench top instrument as well as providing facility for battery charging and data retrieval. The docking station, shown in Figure 2b, allows development of methods by collecting data on representative samples before taking the instrument out to the field. Collection of the calibration data on the same instrument produces the best quality calibration. Furthermore, it allows one instrument to be used for both method development and in field analysis or troubleshooting.
Figure 2
Adhesive bonding provides strong, reliable attachment between similar and dissimilar materials; however, the ultimate strength is often determined by the cleanliness of the two surfaces which are being bonded. To guarantee strong bonding, surfaces should be analytically tested to confirm their cleanliness before application of adhesives. The testing technique should be sensitive to many types of contamination at very low levels. Infrared spectroscopy with a grazing angle sample interface provides a sensitive, universal technique for measuring contamination on metal surfaces; the Exoscan with the grazing angle sample interface allows measurement of large bonding surfaces, making it a truly non-destructive technique.
All infrared spectrometers, including the Exoscan, are sensitive to many different organic and inorganic contaminants. Figure 3 shows spectra obtained with the Exoscan of a thin layer of hydrocarbon oil and a thin layer of silicone, each on an aluminum surface. As can be seen from the spectra, the Exoscan can easily distinguish different types of contaminants, allowing the user to select the proper cleaning procedure to remove the contaminant.
Figure 3
In addition to detecting multiple samples, determination of bond area cleanliness also requires very sensitive detection. Bonding issues are often seen contamination as low as 1 μg/cm2 . Furthermore, the amount of contamination present is important to determine the extent of further cleaning required.
A quantitative calibration to measure the concentration of a silicone contaminant (Frekote™ Mold Release) on aluminum plates was developed. Concentrations from 1.6 to 8.8 μg/cm2 of the silicone on aluminum plates were prepared by spray coating, and then measured with the grazing angle Exoscan. The spectra were collected at 8 cm–1 resolution, collecting 32 scans for a total collection time of 10 seconds. The spectra and a calibration curve using the Si-O band at 1265 cm–1 are shown in Figure 4. The limit of detection for this method was estimated using the band area equivalent to three times (3X) the baseline noise; the LOD was found to be 0.17 μg/cm2 . The LOD could be further increased by using longer measurement times.
Figure 4
The Exoscan was designed to provide easy to understand answers for both skilled and unskilled users. The spectra shown above and are informative for a skilled user to determine the amount and type of signal present, but the successful use of the Exoscan in field applications requires that the results be presented in a simplified form. The calibration curve developed above was entered into the Exoscan's Microlab Mobile software along with predetermined limits allowing the results to be color coded (red, yellow or green). Thus an operator can visually quickly determine whether a particular analyses is within specification. Figure 5 shows an example of this simplified readout which can be used for field use.
Figure 5
High performance, hand-held, portable FT-IR spectrometer provides true non-destructive surface analysis for a wide range of large objects. The Exoscan uses a variety of interchangeable sampling interfaces to accommodate many different sample types; its versatility provides value in both the laboratory and field settings. The system using the grazing angle sample interface has been shown to be effective for determining low concentrations of contaminants on bonding surfaces.
A2 Technologies
14 Commerce Drive, Danbury, CT 06810
Tel. (203) 312-1100, Fax (203) 312-1058
Testing Solutions for Metals and PFAS in Water
January 22nd 2025When it comes to water analysis, it can be challenging for labs to keep up with ever-changing testing regulations while also executing time-efficient, accurate, and risk-mitigating workflows. To ensure the safety of our water, there are a host of national and international regulators such as the US Environmental Protection Agency (EPA), World Health Organization (WHO), and the European Union (EU) that demand stringent testing methods for drinking water and wastewater. Those methods often call for fast implementation and lengthy processes, as well as high sensitivity and reliable instrumentation. This paper explains how your ICP-MS, ICP-OES, and LC-MS-MS workflows can be optimized for compliance with the latest requirements for water testing set by regulations like US EPA methods 200.8, 6010, 6020, and 537.1, along with ISO 17294-2. It will discuss the challenges faced by regulatory labs to meet requirements and present field-proven tips and tricks for simplified implementation and maximized uptime.
Practical Autodilution for ICP-MS and ICP-OES
January 20th 2025Gain insights into improving efficiency and accuracy in elemental analysis through automated dilution technology. Learn about the key capabilities of the Agilent ADS 2 system and its seamless integration with ICP-MS and ICP-OES workflows.
UV-Vis Spectroscopy: Exporting Your Measurement Out of the Instrument
January 20th 2025Optical fibers in ultraviolet-visible (UV-Vis) spectroscopy can enable measurements outside the traditional sample compartment. This paper details the components needed for fiber optic systems, such as couplers and probes, and reviews the performance of Agilent's Cary series instruments. It is crucial to choose the right fiber optic setup for a specific lab’s needs to ensure accurate and efficient measurements.