Spectral measurements to 153 nm can be achieved easily and economically with the high-sensitivity Maya2000 Pro with Extra-Deep-UV Option. Nitrogen purging of the spectrometer helps to mitigate water and oxygen absorption in the vacuum ultraviolet.
Ocean Optics
Spectral measurements to 153 nm can be achieved easily and economically with the high-sensitivity Maya2000 Pro with Extra-Deep-UV Option. Nitrogen purging of the spectrometer helps to mitigate water and oxygen absorption in the vacuum ultraviolet.
A variety of materials have spectral signatures in the vacuum ultraviolet (VUV) spectral region, typically described as the region from 10 nm–200 nm (1,2). VUV spectroscopy is useful for applications ranging from biomedicine to semiconductor testing. Because the spectral response of standard silicon charge-coupled device (CCD) detectors drops off rapidly at wavelengths as long as 400 nm, and because oxygen and water absorb in the VUV spectral band, maintaining sufficient signal outside a vacuum is not possible. Designing vacuum-compatible spectrometers to address these issues can be difficult and expensive. As a result, most commercially available VUV spectrometers require custom configuration and are too expensive and unwieldy for many common applications.
The Maya2000 Pro (3) with Extra-Deep UV Option is poised to change that. The spectrometer features a 101.6-mm focal length optical bench with a compact crossed Czerny-Turner design and a back-thinned CCD detector that offers superior UV and VUV performance. To minimize signal attenuation inside the optical path, the spectrometer bench is purged with nitrogen, achieving a robust signal for VUV experiments at wavelengths as low as 153 nm.
To test performance, we started with a Maya2000 Pro configured with a high resolution, UV-enhanced 2400 g/mm holographic diffraction grating and a 5 µm slit. The back-thinned detector (Hamamatsu S10420-1106) features a peak quantum efficiency of 75%, with UV quantum efficiencies as high as 50%. A custom magnesium fluoride (MgF2) glass window is placed over the detector for improved transmission over VUV wavelengths.
The test sample for the experiment was a VUV deuterium lamp with a VUV-grade optical window, coupled directly to the spectrometer. We enclosed the source and spectrometer and purged the region with standard laboratory-grade nitrogen.
Figure 1
Integration time for the measurement was 50 ms. The resultant emission spectrum shows sharply defined spectral peaks at wavelengths as low as 153 nm, with a strong central peak around 161 nm (see Figure 1). The system operated with a spectral resolution of 0.1 nm and a signal-to-noise ratio of 450:1.
We have also used the system to measure a solution containing carbon, hydrogen/deuterium, nitrogen, oxygen, sulfur, chlorine, bromine, silicon and fluorine chromatographically separated into a column and placed in a vacuum chamber with a helium plasma mixture. VUV spectral features were clearly apparent in the spectrum, suggesting a wide variety of short-wavelength new applications is achievable.
The results demonstrate that the reliable and economical Maya2000 Pro can be easily extended to VUV applications. If you have spectroscopy experiments or applications in the VUV, look no further than the Maya2000 Pro as your short-wavelength, affordable, robust solution.
1. Wikipedia, http://en.wikipedia.org/wiki/Ultraviolet Website covers VUV under larger section on ultraviolet light
2. H. Horspool, F. Lenci, CRC Handbook of Organic Photochemistry and Photobiology (CRC Press, 2003).
3. http://www.oceanoptics.com/Products/maya.asp Web page at OceanOptics.com with full specifications on Maya2000 Pro
Ocean Optics, Inc.
830 Douglas Ave., Dunedin, FL 34698
Phone: +1-727-733-2447; Fax: +1-727-733-3962
Website: www.oceanoptics.com, E-mail: Info@OceanOptics.com
Testing Solutions for Metals and PFAS in Water
January 22nd 2025When it comes to water analysis, it can be challenging for labs to keep up with ever-changing testing regulations while also executing time-efficient, accurate, and risk-mitigating workflows. To ensure the safety of our water, there are a host of national and international regulators such as the US Environmental Protection Agency (EPA), World Health Organization (WHO), and the European Union (EU) that demand stringent testing methods for drinking water and wastewater. Those methods often call for fast implementation and lengthy processes, as well as high sensitivity and reliable instrumentation. This paper explains how your ICP-MS, ICP-OES, and LC-MS-MS workflows can be optimized for compliance with the latest requirements for water testing set by regulations like US EPA methods 200.8, 6010, 6020, and 537.1, along with ISO 17294-2. It will discuss the challenges faced by regulatory labs to meet requirements and present field-proven tips and tricks for simplified implementation and maximized uptime.
Practical Autodilution for ICP-MS and ICP-OES
January 20th 2025Gain insights into improving efficiency and accuracy in elemental analysis through automated dilution technology. Learn about the key capabilities of the Agilent ADS 2 system and its seamless integration with ICP-MS and ICP-OES workflows.
UV-Vis Spectroscopy: Exporting Your Measurement Out of the Instrument
January 20th 2025Optical fibers in ultraviolet-visible (UV-Vis) spectroscopy can enable measurements outside the traditional sample compartment. This paper details the components needed for fiber optic systems, such as couplers and probes, and reviews the performance of Agilent's Cary series instruments. It is crucial to choose the right fiber optic setup for a specific lab’s needs to ensure accurate and efficient measurements.