It is shown that the Retsch CryoMill polymers are ground to a significantly smaller fineness compared to other methods. Especially for inhomogeneous materials, small particles are the key for a high reproducibility.
It is shown that the Retsch CryoMill polymers are ground to a significantly smaller fineness compared to other methods. Especially for inhomogeneous materials, small particles are the key for a high reproducibility.
For many analytical problems in the field of polymer industry XRF (X-Ray Fluorescence) is a very efficient multi element-method for the determination of heavy metals in plastic parts or granulates. In a modern laboratory additionally to the accuracy and reproducibility of the results time security of the laboratory employees are crucial Factors. This is the reason why for more and more applications XRF is preferred because ICP and AAS need time-consuming digestion with strong acids.
In XRF polymers are analysed as powders. To get reliable results, the powder particles should be at least around 500 μm. Smaller particle sizes are advantageous because less disturbing reflections are detected and the material is more homogenous.
For the grinding experiment 3 different mills were used: the Retsch CryoMill, the Retsch MM 400 with CryoKit, and a standard mill for cryo grinding. The sample was a piece of a commercial car tyre. With the MM 400 the sample was pre-cooled in a bath filled with liquid nitrogen (CryoKit) and ground at room temperature. Figure 1 shows the results for the 3 different mills. It can be clearly seen that the result for the MM 400 with CryoKit and the standard cryogenic mill are almost identical. With the Retsch CryoMill a significantly higher fineness can be achieved (>20% better).
Figure 1
In a second experiment the filling of a car seat was ground. Traditionally this is done with a cutting mill, but it is very hard to get particles smaller than 1 mm, because the material gets too hot in the cutting process. For applications like this the CryoMill is very effective — sizes down to 200 μm are easily possible.
For many materials cryogenic grinding is the only method to produce particle sizes below 500 μm which are necessary to obtain reliable results in XRF. The Retsch CryoMill is a perfect tool for grinding polymers which are above the glass point at room temperature. For XRF, grind sizes are obtainable which are significantly lower compared to other mills. Especially for challenging materials like car tyres or fibres the Retsch CryoMill is a perfect tool for XRF sample preparation. Since the Retsch CryoMill works with a liquid nitrogen cycle, the system is both safe and convenient to use.
Retsch GmbH
Rheinische Strasse 36
42781 Haan, Germany
Tel +49 2129 5561-0, Fax +49 2129 8702
Testing Solutions for Metals and PFAS in Water
January 22nd 2025When it comes to water analysis, it can be challenging for labs to keep up with ever-changing testing regulations while also executing time-efficient, accurate, and risk-mitigating workflows. To ensure the safety of our water, there are a host of national and international regulators such as the US Environmental Protection Agency (EPA), World Health Organization (WHO), and the European Union (EU) that demand stringent testing methods for drinking water and wastewater. Those methods often call for fast implementation and lengthy processes, as well as high sensitivity and reliable instrumentation. This paper explains how your ICP-MS, ICP-OES, and LC-MS-MS workflows can be optimized for compliance with the latest requirements for water testing set by regulations like US EPA methods 200.8, 6010, 6020, and 537.1, along with ISO 17294-2. It will discuss the challenges faced by regulatory labs to meet requirements and present field-proven tips and tricks for simplified implementation and maximized uptime.
Practical Autodilution for ICP-MS and ICP-OES
January 20th 2025Gain insights into improving efficiency and accuracy in elemental analysis through automated dilution technology. Learn about the key capabilities of the Agilent ADS 2 system and its seamless integration with ICP-MS and ICP-OES workflows.
UV-Vis Spectroscopy: Exporting Your Measurement Out of the Instrument
January 20th 2025Optical fibers in ultraviolet-visible (UV-Vis) spectroscopy can enable measurements outside the traditional sample compartment. This paper details the components needed for fiber optic systems, such as couplers and probes, and reviews the performance of Agilent's Cary series instruments. It is crucial to choose the right fiber optic setup for a specific lab’s needs to ensure accurate and efficient measurements.