Traditional wet chemistry methods, such as titration or high performance liquid chromatography (HPLC), require complete dissolution of the tablets in suitable solvents, which destroy the samples, create waste, and can be labor intensive and time consuming. Vibrational spectroscopic techniques, notably near-infrared (NIR) absorption and Raman scattering, are nondestructive, fast, and require no consumables. Transmission Raman spectroscopy (TRS) is particularly promising, due to its ability to sample a large portion of the sample’s volume.
QTRam is a compact transmission Raman analyzer designed specifically for content uniformity analysis of pharmaceuticals in solid dosage forms. In this note, we use a model drug, acetaminophen, to demonstrate the capability of QTRam to quantify low concentrations of API in compressed tablets.
Data Collection
QTRam is used to measure transmission Raman spectra of the uncoated tablets. The interrogated sample area is 4 mm in diameter as set by the excitation and collection apertures. BWAnalyst®, the 21 CFR pt 11 compliant software for QTRam, is used to collect all spectra. Each spectrum is an average of 10 scans, and each scan takes 3 s. Two or three spectra were acquired from each tablet sample.
Experimental and Results
Acetaminophen, also known as paracetamol and APAP, is chosen as a model API in this study, due to its availability and low toxicity. We target a hypothetical formulation of 1.5 mg API in a 300 mg tablet, for example, 0.5% APAP w/w. Nine blends were prepared, consisting of APAP, mannitol, silicified microcrystalline cellulose (MCC), croscarmellose, and magnesium stearate (MgSt). The blended powders are compressed into round tablets of 10 mm diameter and 3.0 mm thickness, each weighing roughly 300 mg.
Chemometrics
For quantification, partial least squares (PLS) regression models are built using B&W Tek’s chemometric software BWIQ®. Assuming the API is distributed uniformly in the scale of the interrogated volume, we use the blend concentration as the reference for the individual tablets. Eight tablets are used from each blend, which should reduce the calibration error caused by unideal uniformity by way of averaging.
Conclusion
The QTRam is capable of fast and accurate analysis for content uniformity of low-dose pharmaceutical tablets.
B&W Tek
19 Shea Way, Newark, DE 19713
tel. (302) 368-7824
Website: www.bwtek.com
Measuring Soil Potassium with Near-Infrared Spectroscopy
January 8th 2025Researchers have developed a novel three-step hybrid variable selection strategy, SiPLS-RF(VIM)-IMIV, to enhance the accuracy and efficiency of soil potassium measurement using near-infrared spectroscopy, offering significant advancements for precision agriculture and real-time soil monitoring.
Raman Microscopy for Characterizing Defects in SiC
January 2nd 2025Because there is a different Raman signature for each of the polymorphs as well as the contaminants, Raman microscopy is an ideal tool for analyzing the structure of these materials as well as identifying possible contaminants that would also interfere with performance.