Application Notebook
Moxtek's MAGPRO miniature X-ray source is capable of up to 60 kV 200 µA and down to 4 kV and 1000 µA, and limited to an output power of 12 watts. The 12 W source is meant for use in small bench top systems where small size is valued.
Moxtek's MAGPRO miniature X-ray source is capable of up to 60 kV 200 µA and down to 4 kV and 1000 µA, and limited to an output power of 12 watts. The 12 W source is meant for use in small bench top systems where small size is valued.
The 12 W miniature source has a higher voltage and a much higher emission current than our present miniature X-ray sources, which is an advantage for a number of applications. The initial release of the 12 W source, our MXDPP-50 electronics comes in two versions: the tungsten anode version, which can be used in either XRF or in X-ray imaging, and the Rh anode version, primarily meant for light element XRF detection (Figure 1). The entire X-ray source (tube plus high voltage power supply) weighs about 700 g (25.0 oz) and can be controlled by your choice of USB, digital, or analog control. It also has an interlock connection for X-ray safety.
Figure 1: Image of the 12 W X-ray source.
We will show the advantage of the rhodium anode source running at 4 kV 1000 µA for XRF of cement powders.
Concentrations of sulfur (sulfate), silicon (silicon dioxide), and aluminum (aluminum oxide) are important for both identifying and controlling different blends of cement powder. XRF can be used to measure cement powder elemental concentrations. XRF for the desired "light" elements, with emission lines under about 4 keV is challenging because the characteristic X-rays are absorbed by air, thereby attenuating the signal from the sample. For this experiment, we used a typical XRF setup where the sample is placed at 45° to the tube, and 135° to the detector, and no filters on the 12 W X-ray source. For light element XRF, a rhodium anode is often used because it has L emission lines between 2.7 and 3.0 keV, which are used to excite elements with emission lines below 2.5 keV. For this test the 12 W X-ray source was set to the minimum of 4 kV, which is helpful in eliminating the calcium XRF signal from the cement. The calcium signal is unwanted in this application because it will overwhelm the detection of the other elements of interest. A high emission current X-ray source, such as 1000 µA from the 12 W source, is very helpful in increasing the signal.
Figure 2 shows an XRF spectrum from a Portland cement sample with the 12 W source running at 1000 µA and 200 µA on a rhodium anode tube. The signal level is increased by a bit over fivefold for these light elements, which can be a huge advantage in any application which detection of light elements is needed.
Figure 2: Running at 1000 µA, the 12 W source produces much more signal from the desired elements in the cement powder than running at 200 µA.
The new 12 W source is meant for use in small benchtop systems where small size is valued. It has as increased range up to 60 kV at high voltage and increased current to 1000 µA at low voltages; both of which are an advantage in XRF and imaging applications.
Moxtek
452 West 1260 North, Orem, UT 84057
tel. (800) 758-3110, email: info@moxtek.com
Website: www.moxtek.com
Atomic Perspectives: Highlights from Recent Columns
March 3rd 2025“Atomic Perspectives,” provides tutorials and updates on new analytical atomic spectroscopy techniques in a broad range of applications, including environmental analysis, food and beverage analysis, and space exploration, to name a few. Here, we present a compilation of some of the most popular columns.
Pittcon 2025: Highlighting Talks on Atomic Spectroscopy
February 26th 2025At Pittcon this year, there will be numerous sessions dedicated to spotlighting the latest research that uses atomic spectroscopy or elemental analysis techniques. We highlight some of these talks below that might pique the interest of spectroscopists and researchers attending the conference this year.
The 2025 Emerging Leader in Atomic Spectroscopy Award
February 15th 2025Benjamin T. Manard has won the 2025 Emerging Leader in Atomic Spectroscopy Award for his pioneering research in nuclear material characterization and isotope ratio analysis, with expertise in advanced atomic spectrometry techniques such as inductively coupled plasma optical emission spectroscopy (ICP-OES), inductively coupled plasma mass spectrometry (ICP-MS), and laser ablation.
Applications of Micro X-Ray Fluorescence Spectroscopy in Food and Agricultural Products
January 25th 2025In recent years, advances in X-ray optics and detectors have enabled the commercialization of laboratory μXRF spectrometers with spot sizes of ~3 to 30 μm that are suitable for routine imaging of element localization, which was previously only available with scanning electron microscopy (SEM-EDS). This new technique opens a variety of new μXRF applications in the food and agricultural sciences, which have the potential to provide researchers with valuable data that can enhance food safety, improve product consistency, and refine our understanding of the mechanisms of elemental uptake and homeostasis in agricultural crops. This month’s column takes a more detailed look at some of those application areas.
Best of the Week: Seed Vigor, Flower Classification, Emerging Leader in Atomic Spectroscopy
January 10th 2025Top articles published this week include two peer-reviewed articles that explore optical detection technology for seed vigor and classifying flowers, as well as a profile on Benjamin Manard, who was recognized as the winner of the 2025 Emerging Leader in Atomic Spectroscopy.