Application Notebook
The United States Pharmacopeia (USP) is in the process of developing new USP chapters and for the measurement of inorganic contaminants in pharmaceutical samples. While samples can be dissolved in aqueous or organic solvents, the large majority will require digestion, and closed vessel microwave digestion is stipulated by USP. Single reaction chamber (SRC) microwave digestion is a new type of closed vessel digestion that was developed by Milestone and offers many advantages over closed vessel digestion-particularly for pharmaceutical sample types.
Single reaction chamber microwave digestion offers significant benefits over traditional microwave digestion for pharmaceutical samples: higher sample throughput, reduced labor and consumables costs, simplified workflow, and superior digestion quality.
The United States Pharmacopeia (USP) is in the process of developing new USP chapters <232> and <233> for the measurement of inorganic contaminants in pharmaceutical samples. While samples can be dissolved in aqueous or organic solvents, the large majority will require digestion, and closed vessel microwave digestion is stipulated by USP. Single reaction chamber (SRC) microwave digestion is a new type of closed vessel digestion that was developed by Milestone and offers many advantages over closed vessel digestion-particularly for pharmaceutical sample types.
The Milestone UltraWAVE can digest up to 15 samples simultaneously, at up to 300 °C and 199 bar pressure. The high temperature and pressure capabilities enable the complete digestion of virtually every sample type. Samples are weighed into vials and placed in the chamber: any combination of sample types can be digested together-greatly simplifying the sample prep workflow. Raw materials, excipients, API, and final product can all be digested together in the same run. The Milestone UltraWAVE uses disposable glass vials, eliminating vessel assembly/disassembly and vessel cleaning, which significantly reduces labor and consumables costs. Most pharmaceutical sample types have high organic content, including oils and fats that are difficult to digest and generate CO2, which increases pressure in the reaction vessel. The high pressure capability of the UltraWAVE enables it to digest entire fish oil gelcaps-not possible with traditional closed vessel microwave.
Table I: St. John's wort
The UltraWAVE can digest samples in as little as 3 mL acid, which minimizes reagent blank levels-important since ICP-MS will be used predominantly for USP <232> and <233> when implemented.
Table II: Fish oil gelcaps
The SRC chamber is pre-pressurized with N2 prior to the start of the run. This prevents sample boiling, eliminating cross-contamination and loss of volatiles.
Table III: Quality control summary (fish oil gelcaps)
Due to the UltraWAVE's higher sample capacity, use of disposable vials, and faster cool down time, sample throughput is 2–3× higher than closed vessel digestion. Its lower consumables costs, simple operation, and superior digestion quality also make the Milestone UltraWAVE the perfect choice for pharmaceutical sample prep.
The data above shows excellent DLs and recovery-even for the volatile element mercury-no cross-contamination with the UltraWAVE.
Milestone, Inc.
25 Controls Drive, Shelton, CT 06484
tel. (866) 995-5100, fax (203) 925-4241
Website: www.milestonesci.com
Atomic Perspectives: Highlights from Recent Columns
March 3rd 2025“Atomic Perspectives,” provides tutorials and updates on new analytical atomic spectroscopy techniques in a broad range of applications, including environmental analysis, food and beverage analysis, and space exploration, to name a few. Here, we present a compilation of some of the most popular columns.
Pittcon 2025: Highlighting Talks on Atomic Spectroscopy
February 26th 2025At Pittcon this year, there will be numerous sessions dedicated to spotlighting the latest research that uses atomic spectroscopy or elemental analysis techniques. We highlight some of these talks below that might pique the interest of spectroscopists and researchers attending the conference this year.
The 2025 Emerging Leader in Atomic Spectroscopy Award
February 15th 2025Benjamin T. Manard has won the 2025 Emerging Leader in Atomic Spectroscopy Award for his pioneering research in nuclear material characterization and isotope ratio analysis, with expertise in advanced atomic spectrometry techniques such as inductively coupled plasma optical emission spectroscopy (ICP-OES), inductively coupled plasma mass spectrometry (ICP-MS), and laser ablation.
Applications of Micro X-Ray Fluorescence Spectroscopy in Food and Agricultural Products
January 25th 2025In recent years, advances in X-ray optics and detectors have enabled the commercialization of laboratory μXRF spectrometers with spot sizes of ~3 to 30 μm that are suitable for routine imaging of element localization, which was previously only available with scanning electron microscopy (SEM-EDS). This new technique opens a variety of new μXRF applications in the food and agricultural sciences, which have the potential to provide researchers with valuable data that can enhance food safety, improve product consistency, and refine our understanding of the mechanisms of elemental uptake and homeostasis in agricultural crops. This month’s column takes a more detailed look at some of those application areas.