Application Notebook
The permitted sulfur limit in fuel oils has been decreased to 10 ppm in many regions.
The permitted sulfur limit in fuel oils has been decreased to 10 ppm in many regions. This application note demonstrates that the performance of the Rigaku Micro-Z ULS meets the requirement of ASTM D2622-10, in the analysis of ultra-low sulfur in automotive fuels.
Recent developments in ultra-low sulfur (ULS) fuel have improved fuel efficiency and created cleaner exhaust gas. Globally, the permitted sulfur limit in fuel oils has been decreased to 10 ppm in many countries and regions.
For compliance verification, X-ray fluorescence (XRF) spectrometry is the definitive analysis tool for use at distribution terminals and refineries, as well as mobile or stationary testing laboratories. In recent years, there has been an increasing need for an instrument which does not require the use of helium gas, for instances when acquisition or delivery of helium to the analysis site is difficult.
The Micro-Z ULS (ultra-low sulfur) is a newly developed sulfur analyzer which does not require helium gas in operation. This application note demonstrates that Micro-Z ULS can meet the requirements of ASTM D2622-10.
Instrument
The Micro-Z ULS is a benchtop wavelength dispersive X-ray fluorescence (WDXRF) spectrometer with fixed optics optimized for sulfur analysis. The atmosphere in the optics path of the Micro-Z ULS is a vacuum, so that helium gas is not required. The spectrometer is designed to minimize the restrictions in installation such as the need for cooling water, a special power supply, and installation space. The Micro-Z ULS is equipped with an air-cooled 40 W Cr-target X-ray tube and a doubly curved RX-9 analyzing crystal optimized for low concentration sulfur.
The Micro-Z ULS has a built-in operation panel, which enables simple and easy operation for daily analysis. The counting time was 300 s for peak and background measurements in this experiment.
Sample Preparation
Approximately, 4 mL of each sample was poured into a liquid cell with sample film of 2.5 µm Mylar® film (Chemplex® 100).
Standard and Calibration
"Number 2 diesel fuel" standards and isooctane-based standards provided by VHG Labs, Inc. were used for calibrations of diesel fuel and gasoline respectively.
Figure 1: Calibration curve for each material of Micro-Z ULS; (a) diesel fuel, (b) isooctane (for gasoline).
Analysis Results
Repeatability tests were carried out using a representative sample for each material. For each sample, two aliquots were prepared and analyzed using the calibration Figure 1; this process was repeated 20 times. The term "r" represents "repeatability" defined by the following formula in ASTM D2622-10 (1):
The difference between successive test results obtained by the same operator with the same apparatus under constant operation conditions on identical test material would exceed the values calculated by equation 1 only in one case in every 20.
The test results, where the maximum value for each fuel is smaller than r, prove that the performance of the Micro-Z ULS meets the requirement of ASTM D2622-10 for diesel fuel and gasoline.
Low concentration sulfur in petroleum-based fuel can be routinely analyzed with high accuracy and precision on the Micro-Z ULS, a benchtop WDXRF spectrometer with fixed optics, which does not require He gas, external water for cooling the X-ray tube, or special power supply. This application note demonstrates that the performance of the Micro-Z ULS meets the requirement of ASTM D2622-10, which has become strict in the recent versions of ASTM D2622.
(1) ASTM D2622-10(2010), "Standard Test Method for Sulfur in Petroleum Products by Wavelength Dispersive X-ray Fluorescence Spectrometry," ASTM International, p. 12.
Rigaku Corporation
9009 New Trails Drive, The Woodlands TX 77381
tel. (281) 362-2300, fax (281) 364-3628
Website: www.rigaku.com
Poster Presentation - RISE imaging of various phases of SiC in sintered silicon-carbide ceramics
November 12th 2024In this poster presentation applications manager Ute Schmidt discusses the use of correlative Raman-SEM (RISE Microscopy) imaging for analyzing silicon-carbide (SiC) ceramics. The main focus of the work is investigating the distribution of sub-micron structured polytypes of SiC grains on and below the surface.
FT-IR gas analysis of coal-to-ethylene glycol process
October 31st 2024Coal-to-ethylene glycol is the process by which ethylene glycol is synthesized from coal instead of traditional methods using petroleum as the raw material. This study demonstrates that Fourier Transform Infrared Spectroscopy (FT-IR) can be a reliable alternative to simultaneously measure methyl nitrite and other process gases (such as CO and NO) using a single ABB analyzer.