Photon etc. has designed two narrowband tunable filters for resonance Raman spectroscopy.
Marc Verhaegen and Sébastien Blais-Ouellette, Photon etc.
Photon etc. has designed two narrowband tunable filters for resonance Raman spectroscopy. These matched tunable passband and notch filters, based on thick volume Bragg gratings, exhibit a bandwidth less than 10 cm-1 and cover a wavelength range of several hundreds of nanometers.
Raman spectroscopy (RS) is a powerful tool to study vibrational, optical, and electronic properties of materials in a nondestructive manner. Raman signals are typically orders of magnitude lower than the intensity of the excitation laser line (1). However, it is possible to significantly increase Raman signals by choosing an excitation wavelength that corresponds to an optical transition of the material under investigation.
Resonant Raman Spectroscopy (RRS) provides a unique tool to characterize the diameter and chirality distribution of a mixed population of carbon nanotubes (CNT). RRS is also a powerful method to monitor in-situ the CNT properties during growth. Thus, it can serve as a diagnosis tool in order to achieve better control of the CNT production.
Using volume Bragg grating (VBG) technology, Photon etc. has developed, specifically for RRS, two types of ultra narrow band tunable filters: a laser line filter and a notch filter.
The Laser Line Tunable Filter (LLTF) is installed just after the Tunable Ti:Saph laser (Figure 1) and blocks the unwanted fluorescence produced by the laser, leaving the excitation laser line untouched. Two steering mirrors (M1 and BS) send the laser line into a standard microscope where the laser beam is focused on the sample of interest.
Figure 1: Experimental setup for Raman characterization.
The second filter, a Tunable Top Notch™ Filter (TTNF), is installed after the microscope. The TTNF blocks the Rayleigh scattering coming from the material, leaving the Raman signal untouched down to 50 cm-1 with a throughput up to 80%.
The tunable filters are controlled by a computer via USB links, allowing fast wavelength selection.
Stokes and anti-Stokes Raman spectra of single-walled carbon nanotube (SWNT) powder (Figure 2) were measured within less than 1 h using a standard spectrometer (2). Each peak corresponds to a radial breathing mode (RBM), and the center frequency of a given peak is inversely proportional to the nanotube diameter of a given population. Several populations of nanotubes with different diameters can therefore be readily observed and effectively characterized.
Figure 2: Raman spectra of CNT using Ti:Saph laser and TTNF.
(1) J. R. Ferraro, K. Nakamoto, and C. W. Brown, Introductory Raman Spectroscopy, 2nd Edition (2003).
(2) M. Paillet, F. Meunier, M. Verhaegen, S. Blais-Ouellette, and R. Martel, Rev. Sci. Instrum.81, 1 (2010).
Photon etc.
5795, avenue de Gaspé, #222, Montreal, Quebec, H2S 2X3
tel. (514) 385-9555; fax (514) 279-5493
Website: www.photonetc.com
The Rise of Smart Skin Using AI-Powered SERS Wearable Sensors for Real-Time Health Monitoring
May 5th 2025A new comprehensive review explores how wearable plasmonic sensors using surface-enhanced Raman spectroscopy (SERS) are changing the landscape for non-invasive health monitoring. By combining nanotechnology, AI, and real-time spectroscopy analysis to detect critical biomarkers in human sweat, this integration of nanomaterials, flexible electronics, and AI is changing how we monitor health and disease in real-time.
Nanometer-Scale Studies Using Tip Enhanced Raman Spectroscopy
February 8th 2013Volker Deckert, the winner of the 2013 Charles Mann Award, is advancing the use of tip enhanced Raman spectroscopy (TERS) to push the lateral resolution of vibrational spectroscopy well below the Abbe limit, to achieve single-molecule sensitivity. Because the tip can be moved with sub-nanometer precision, structural information with unmatched spatial resolution can be achieved without the need of specific labels.
AI-Powered SERS Spectroscopy Breakthrough Boosts Safety of Medicinal Food Products
April 16th 2025A new deep learning-enhanced spectroscopic platform—SERSome—developed by researchers in China and Finland, identifies medicinal and edible homologs (MEHs) with 98% accuracy. This innovation could revolutionize safety and quality control in the growing MEH market.