SPEC PerkinElmer 10.13
|Articles|September 1, 2008

Application Notebook

  • Application Notebook-09-01-2008
  • Volume 0
  • Issue 0

Raman Quantitative Calibration Stability for a Sample Mixture

The simplicity of Raman spectroscopy provides an ideal method for the quantitative determination of liquid mixtures.

The simplicity of Raman spectroscopy provides an ideal method for the quantitative determination of liquid mixtures.

The use of Raman spectroscopy for routine quantitative analysis is desirable due to the simple sample handling requirements for sample analysis. Due to the many variables associated with Raman spectroscopy, specifically, laser power, Raman cross-section, wavelength and intensity stability, Raman is rarely used for quantitative analysis. Compared to Raman analysis, there are other methods that can provide a repeatable quantitative prediction of unknowns, but these methods can require extensive sample preparation or extended time for analysis. To demonstrate the capabilities of Raman spectroscopy for quantitative analysis, a two-component mixture of liquid samples was analyzed using a Raman spectrometer and the results reported.

Experimental

Several mixtures of benzonitrile dissolved in ethanol were prepared at 0.1, 0.3, 0.5, 1.0, 3.0 and 5.0% concentration as standards for development of the multi-component calibration model including a 0.2% mixture to be used as an 'unknown' for analysis using the calibration model. A Jasco NRS-3100 Raman spectrometer was used to collect triplicate spectra of the standards using 532 nm excitation and a 60 s integration of the CCD with two accumulations to obtain spectra with a spectral resolution of slightly less than 4 cm–1 resolution. Samples were contained within 2 mL HPLC sample vials and a macro accessory used for excitation/collection of the sample volumes. No further correction of the Raman spectra was performed prior to development of the calibration model or analysis of 'unknown' spectra.

Results

The various standard spectra as well as an ethanol 'blank' were used to develop the quantitative calibration model using Jasco's Spectra Manager Quantitative Analysis package. A peak ratio method using the 2234 cm–1 C-N stretch of benzonitrile vs. the 1455 cm–1 C-H bend of the ethanol was used for the development of the linear calibration curve. Initial results based on the initial triplicate spectra of each standard provided a calibration curve with a fit of 0.9998 and an R2 of 0.9996, indicating an excellent fit of the data. To evaluate the repeatability of the calibration method, additional spectral data sets of all standards were collected again after one and two weeks, providing a total of at least nine spectra of each standard for calculation of the calibration curve. The final calibration curve also provided a linear calibration curve with very little scatter, offering a fit of 0.9997 and an R2 (residual) of 0.9995 (Figure 1). As well, the analysis of the 'unknown' sample that was prepared indicated errors of less than 0.5% for the replicate analyses conducted over the two-week period.

Figure 1: Calibration model for quantitative standard mixtures, including 2nd and 3rd week standards analysis.

Conclusions

The results demonstrate that a reliable calibration model can be developed from Raman spectra of a mixture without further correction. The results demonstrate a correlation of 0.999 or better for the quantitative component over a range of concentrations from approximately 0.1 to 5%. Further, the Raman method provides repeatable, reproducible results and proves to be stable for extended periods.

Jasco, Inc.

8649 Commerce Dr., Easton, MD 21610

Tel. 800-333-5272; Fax 410-822-7526

www.jascoinc.com


Related to this article

FACSS 2026 Award Interviews ©  Erin -chronicles-stock.adobe.com
Eight FACSS award winners at SciX 2026, One LIBS trailblazer. Forty-eight questions. And not one of them is a softball. Award season in spectroscopy usually means polite applause, a plaque, and a photo. We're not completely interested only in the award sessions. The eight scientists honored at SciX 2026 in Sparks, Nevada, along with LIBS researcher Alessandro De Giacomo, are pushing Raman into operating rooms, flying LIBS on drones, reading chemistry off Mars, and tracing toxic metals downwind of industrial sites. Their work makes big claims. In the coming days, Spectroscopy will sit down with eight of these researchers and ask whether those claims hold up. The interviews that follow won't just celebrate. They'll press on the gaps between simulation and experiment, between the lab bench and the clinic, and between a clever paper and an instrument someone will actually buy and use.
Spectral Frontiers: Past the Diffraction Limit (FT-IR Spectroscopy)
Fourier-transform infrared (FT-IR) spectroscopy, long the workhorse of molecular fingerprinting, is being reinvented by photothermal detection schemes and quantum cascade laser (QCL) sources that push spatial resolution and analysis speed far beyond classical dispersive and interferometric limits. New instrumentation such as optical photothermal infrared (O-PTIR) microscopy and laser direct infrared (LDIR) chemical imaging is enabling sub-micron, label-free chemical mapping of biological, environmental, and pharmaceutical samples. This article reviews the latest FT-IR-adjacent technologies, recent patents, and research driving the field, featuring perspectives from the scientists and instrument developers advancing the technique.