Wavelength-Scanned Cavity Ring Down Spectroscopy combines the real-time speed and turnkey simplicity of optical spectroscopy with the precision (<0.1 δ18O [‰], < 0.5 δD [‰]) previously only available from complex IRMS systems.
Wavelength-Scanned Cavity Ring Down Spectroscopy combines the real-time speed and turnkey simplicity of optical spectroscopy with the precision (<0.1 δ18O [‰], < 0.5 δD [‰]) previously only available from complex IRMS systems.
The fractionation of the isotopes of water that occurs during the hydrologic cycle (1) enables studies of many different processes such as climate change, ice and water flow, aquifer dynamics and landfill leaks. In addition, isotopically labeled water is used to measure metabolism processes including energy expenditure in humans (2).
WS-CRDS systems are increasingly used to measure isotopic ratios of 13C/12C, D/H, 18O/16O in atmospheric gas phase measurements (3). This application note demonstrates how a benchtop WS-CRDS instrument can simultaneously measure both δ18O and δD for liquid water samples quickly and easily without the need for special sample preparation or extensive operator training.
This study used a benchtop WS-CRDS water isotope analyzer (Picarro L1102-i) including an autosampler and vaporizer. The analyzer is equipped with a thermally controlled optical cavity with an effective path length of 20 kms. In addition, an onboard wavelength monitor ensures that the absorption lines unique to H216O, H218O, and HD16O can be scanned repeatedly and precisely, providing both δ18O and δD data in one fast measurement. The robotic sampler is configured to access a rack of liquid water samples in standard 2 ml vials. Liquid water is injected into the vaporizer and the ensuing water vapor is passed into the WS-CRDS cavity. The instrument method can be adjusted to balance throughput with precision and in this study the method is timed at 9 min to maximize precision. Experimental set-up can be easily accomplished through a .csv or similar file which presents the method and other parameters for the automated run. Data is output in a format suitable for standard processing techniques.
Figure 1 shows data obtained from four different groundwater locations spanning a range of both δ18O and δD. Samples were filtered to remove particulates prior to the run. Each point represents a single 2 μl injection and 25 injections were made from each location. The first injection from each sample contained memory effects and was discarded. However, all injections were run contiguously without additional forms of purging between the different samples. As shown, the standard deviation for each location is remarkably small, especially given the change in environmental conditions over a non-stop 17 h of instrument operation. This level of repeatability with no drift confirms that a large batch of samples can be run without the need for re-calibration with reference samples, in sharp contrast to IRMS methods.
Figure 1
This demonstration of easy access to precise measurement of δ18O and δD should further accelerate and diversify the use of stable isotopes, in fields as diverse as water cycle studies, bird migration mapping, and metabolic activity modeling and optimization, to name just a few.
(1) N.L. Ingraham, in Isotope Tracers in Catchment Hydrology, C. Kendall and J.J. McDonnell, Eds. (Elsevier Science B.V., Amsterdam, 1998), pp. 87–115.
(2) D.A. Schoeller, et al., Int. J. Obesity 25(8), 1240–1245 (2001).
(3) E.R. Crosson, et al, Isotopes Environ. Health Stud. 42(1), 21–35 (2006).
Picarro, Inc.
480 Oakmead Parkway
Sunnyvale, CA 94085
Tel. (408) 962-3900, Fax (408) 962-3200
962-3200
Testing Solutions for Metals and PFAS in Water
January 22nd 2025When it comes to water analysis, it can be challenging for labs to keep up with ever-changing testing regulations while also executing time-efficient, accurate, and risk-mitigating workflows. To ensure the safety of our water, there are a host of national and international regulators such as the US Environmental Protection Agency (EPA), World Health Organization (WHO), and the European Union (EU) that demand stringent testing methods for drinking water and wastewater. Those methods often call for fast implementation and lengthy processes, as well as high sensitivity and reliable instrumentation. This paper explains how your ICP-MS, ICP-OES, and LC-MS-MS workflows can be optimized for compliance with the latest requirements for water testing set by regulations like US EPA methods 200.8, 6010, 6020, and 537.1, along with ISO 17294-2. It will discuss the challenges faced by regulatory labs to meet requirements and present field-proven tips and tricks for simplified implementation and maximized uptime.
Practical Autodilution for ICP-MS and ICP-OES
January 20th 2025Gain insights into improving efficiency and accuracy in elemental analysis through automated dilution technology. Learn about the key capabilities of the Agilent ADS 2 system and its seamless integration with ICP-MS and ICP-OES workflows.
UV-Vis Spectroscopy: Exporting Your Measurement Out of the Instrument
January 20th 2025Optical fibers in ultraviolet-visible (UV-Vis) spectroscopy can enable measurements outside the traditional sample compartment. This paper details the components needed for fiber optic systems, such as couplers and probes, and reviews the performance of Agilent's Cary series instruments. It is crucial to choose the right fiber optic setup for a specific lab’s needs to ensure accurate and efficient measurements.